The AngE Pro 8 is an 8-channel puls oscillography device with additional ECG-trigger. It is the extension of our basis device, the
AngE Angio Experience 4-channel.

The AngE Pro 8 offers measurement methods for the simultaneous and segmental measuring. First of all, this leads to a better comparison of the vessels in the extremities and an improved Doppler-Index calculation.

The puls waves are recorded at different cuff pressures. The mean arterial pressure is determined by the maximum amplitude. The comparison of the mean arterial pressure leads to a result similar to the ABI.
Additionally, the AngE has a built-in ECG-Trigger which works simultaneous to the cuff-measurements, allowing the measurement and calculation of the puls wave runtime and puls wave velocity.

The alternative solution to ABI

AngE Pro 8 Cuffs

These days, even in our health care system, cost effective operation is a necessity.
The AngE Pro 8 has been developed to obtain the diagnosis of a peripheral arterial occlusive disease quickly, reliably and with cost effectiveness, without stress for the patient.
Therefore, the classic, more time consuming and hard to document ABI measurement is a relic from the past.
Using the oscillometric principle developed by Gesenius and Keller, the pulse waves are recorded at different cuff pressures. The mean arterial pressure is determined by maximum amplitude. The comparison of the mean arterial pressure leads to a result similar to the ABI.

AngE Finger/Toe-Cuffs



The system is able to record the pulse waves at fingers and toes, thigh and lower leg as well as ankle and wrist at 8 channels. Due to the simultaneous measurement, there is no time difference between the measurements on the single extremities as with the ABI measurement. This delay between the individual determinations of blood pressure values leads to an inaccuracy which is avoided by the AngE. In addition the simultaneous recording allows the comparison of time values between the limbs.


Patients suffering from diabetes often have very calcified vessels (mediasclerosis) which can not be compressed. Therefore, the ABI results are unusable.

Using the finger and toe measurement of the AngE, reliable results can be obtained, because the mediasclerosis affects the toes only in less than 1% of such patients.
Measuring this way, even with patients suffering from mediasclerosis, a fast and consistent diagnosis is possible. This early detection of pAOD avoids lots of complications for the patient.

Pulswave

Puls Quality

One of the AngE Pro 8's best improvements is the better displayed puls wave. Using better pressure sensors and an elaborated filter algorithm, the puls wave is more sensitive and detailed than before. This lets the examiner recognize even smallest spikes.

The AngE Pro 8 provides a multitude of additional parameters compared to the ABI measurement. Where the ABI only provides pressure values of the limbs, AngE records the puls wave which allows obtaining a detailed puls wave analysis. Parameters such as rise time, amplitude and the optical appearance of the puls wave can be evaluated.

The calculation of RI(reflection index) and SI(stiffness index) allow the identification of patients with a high vascular risk by the inclusion of the dicrotic wave into the measurement. With medicamentous provocation, the endothelial function can be examined.

ECG

ECG Trigger

With the recording of the ECG signal it's possible to use the R-wave to obtain a trigger point, thus allowing the measurement and calculation of pulse wave runtime and pulse wave velocity. This parameters increase in their importance according to the evaluation of the state of the vessels e.g. in clinical monitoring of patients suffering from hypertension.

The AngE Pro 8 is shipped with two electrode clamps in order to minimize the effort involved in the measurement. The electrodes just need to be placed onto the patients wrists.

The advantages of the oscillographic principle following the Gesenius method

AngE Angio Experience

  • Examaination independent method 
  • Quich and trouble free, results are available within 3-4 minutes 
  • Fully automatic index calculation 
  • Highest possible accuracy on the detection of peripheral arterial occlusion diseases. Dedicates for the detection of media sclerosis and Subclavia occlusion 
  • Documenation of side differences
  • Subclavian desease recognition
  • Automatic calculation of the peak time and the quotient between rise and fall time with documentaion of the normal value 
  • Windows-based Analysis Software with documentaion 
  • Ideal for process controls of the peripheral artterial occlusion diseases (with diabetic patients, too) 
  • Economical procedure